3.3 Incident Light Operation: Epi-Fluorescence
For microscopes with fluorescence option installed only.
For viewing of transparent objects using incident-light fluorescence, we recommend to
first make your initial adjustments using brightfield with transmitted light.
1. First, switch on the power supply to the Nobska lamphouse and allow the lamp to
achieve operating temperature. (Operating temperature is achieved when the power
supply settles on the proper operating voltage. See your Nobska Operations Manual for
more details.)
2. Open the light stop leading to the
lamphouse.
3. Move filter block into the light
beam path.
4. Remove the objective currently in
the light path.
5. Place a sheet of white paper on the
specimen stage.
6. Observe if the arc image is clearly
projected and centered in the light spot as shown in Figure X. (If not, proceed to the
section in the Nobska Lamphouse Manual describing Mercury Lamp arc position
adjustment.)
7. Remove the paper and re-install the objective.
8. Position the specimen on the stage.
9. Using a low power objective, check to see if the image is illuminated evenly. If not,
adjust the settings on the Nobska Lamphouse if needed.
Be careful not to project the arc reflection onto the arc electrodes. This will cause an
over-heating condition followed by an explosion of the mercury lamp.
The two electrodes of the arc can be seen in the extension of the plane of the discharge
arc. Refer to the Nobska Lamphouse Manual for further information and adjustment
instructions.